Remote Display True-rms Clamp Meter
Calibration Adjustment
15
Calibration Adjustment
Calibration Password Operation Procedure
A password is required to access the Calibration adjustment section of the Product. This
section explains how this functionality works. To enter a password:
1. Push and turn the rotary switch to . The display shows “CAL”.
2. Push to enter the password entry mode. The display shows “????”.
3. Enter the 4-digit password. For this step, the keys are mapped to the corresponding
digits:
- 1
- 2
- 3
- 4
- 5
• If you enter a incorrect password, push or . The display shows “????”
and you can re-enter the password.
• If you enter the correct password, push to continue calibration (step 5) or
push to change the password. See Change the Password.
4. Push to enter calibration mode.
5. Once the Product has entered calibration mode, push to show the calibration
step number. Push to confirm a calibration value.