Remote Display True-rms Clamp Meter
Performance Tests
13
Performance tests for the iFlex are in Table 4. Connect the 5520A A ac output to the 50-
turn coil. Figures 6 and 7 show the performance test connections for the iFlex. For the
iFlex simulated tests you need the leads and probes listed in the required equipment list.
They are used to connect the calibrator output to the iFlex input. Calibrator Output HI
goes to the iFlex 2 mm jack (on the far left) and the Calibrator Output LO goes to the
black COM jack. See the notes in Table 4.
50-Turn Current Coil
Conductor
ghn13.eps
Figure 6. Performance Test Connections for the iFlex