EasyManuals Logo

Fluke 45 User Manual

Fluke 45
122 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #100 background imageLoading...
Page #100 background image
45
Users Manual
6-6
Table 6-2. Performance Tests for Volts, Diode Test, Ohms, and Frequency, Functions (cont.)
Display
Function Range (Rate)
Input Level
Frequency
Min Max
Using decades of 3:
300
3 k
30 k
300 k
3 M
30 M*
300 M*
Short
300
short
3 k
30 k
300 k
3 M
30 M
300 M
Using decades of 1.9:
300
3 k
30 k
300 k
3 M
30 M*
300 M*
Short
190
short
1.9 k
19 k
190 k
1.9 M
19 M
190 M
Using decades of 1:
300
3 k
3 k
30 k
300 k
3 M
30 M*
300 M*
Short
100
short
1 k
10 k
100 k
1 M
10 M
100 M
0.00
299.83
0.0000
2.9983
29.983
299.83
2.9980
29.922
294.0
0.00
189 88
0.0000
1.8988
18.988
189.88
1.8987
18.949
186.2
0.00
99.93
0.0000
0.9993
9.993
99.93
0.9992
9.972
98.0
0.04
300.19
0.0002
3.0017
30.017
300.17
3.0020
30.078
306.0
0.04
190.14
0.0002
1.9012
19.012
190.12
1.9013
19.051
193.8
0.04
100.09
0.0002
1.0007
10.007
100.07
1.0008
10.028
102.0
FREQ 10/100 kHz 0.1 V to 10 V 10 kHz 9.9949 10.006
* Optional test points that can be used if standards are available.

Table of Contents

Other manuals for Fluke 45

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 45 and is the answer not in the manual?

Fluke 45 Specifications

General IconGeneral
Diode TestYes
Display Digits5 digits
InterfaceRS-232
Frequency Range5Hz to 500kHz
Continuity< 25 Ω
Basic DC Accuracy0.05%
Power Supply115/230 V AC, 50/60 Hz