DC Voltage
Resolution AccuracyRange
Slow Medium Fast (6 Months) (1 Year)
300 mV
3 V
30 V
300 V
1000 V
10 µV
100 µV
1
mV
10 mV
100 mV
100 µV
1 mV
10 mV
100 mV
1 V
002 % + 2
0.02 % + 2
0.02 % + 2
0.02 % + 2
0.02 % + 2
0.025 % + 2
0.025 % + 2
0.025 % + 2
0.025 % + 2
0.025 % + 2
100 mV
1000 mV
10 V
100 V
1000 V
1 µV
10 µV
100 µV
1 mV
10 mV
0.02 % + 6
0.02 % + 6
0.02 % + 6
0.02 % + 6
0.02 % + 6
0.025 % + 6
0.025 % + 6
0.025 % + 6
0.025 % + 6
0.025 % + 6
Input Impedance
10 Me in parallel with <100 pF
Note
In the dual display mode, when the volts ac and volts dc functions are
selected, the 10 M
dc input divider is in parallel with the 1 M
ac
divider.
Normal Mode Rejection Ratio
>80 dB at 50 Hz or 60 Hz, slow and medium rates
>54 dB for frequencies between 50-440 Hz, slow and medium rates
>60 dB at 50 Hz, fast rate (Note: Fast rate has no filtering)
Maximum Allowable AC Voltage While Measuring DC Voltage or (AC + DC)
Voltages
Peak Normal Mode SignalRange Max Allowable Peak AC
Voltage
NMRR* >80 dB NMRR >60 dB
300 mV
3 V
30 V
300 V
1000 V
100 mV
1000 mV
10 V
100 V
1000 V
15 V
15 V
1000 V
1000 V
1000 V
15 V
15 V
50 V
50 V
200 V
15 V
15 V
300 V
300 V
1000 V
* NMRR is the Normal Mode Rejection Ratio
Normal Mode Rejection Ratio at 50 Hz or 60 Hz ±0.1 %
Common Mode Rejection Ratio
>90 dB at do, 50 or 60 Hz, (1 ke unbalanced, medium and slow rates)

Table of Contents

Other manuals for Fluke 45

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 45 and is the answer not in the manual?

Fluke 45 Specifications

General IconGeneral
Diode TestYes
Display Digits5 digits
InterfaceRS-232
Frequency Range5Hz to 500kHz
Continuity< 25 Ω
Basic DC Accuracy0.05%
Power Supply115/230 V AC, 50/60 Hz