Calibrator
General Specifications
5
• Singapore: +65-738-5655
• Anywhere in the world: +1-425-446-5500
Or, visit Fluke's website at www.fluke.com
.
To register your product, visit http://register.fluke.com
.
To view, print, or download the latest manual supplement, visit
http://us.fluke.com/usen/support/manuals
.
General Specifications
All specifications are valid after a warm-up period of 30 minutes, or twice the time since last warmed up, to a maximum of
30 minutes. For example, if the 5080A has been turned off for 5 minutes, the warm-up period is 10 minutes.
All specifications apply for the temperature and time period indicated. For temperatures outside of tcal ±5 °C (tcal is the
ambient temperature when the 5080A was calibrated), the temperature coefficient as stated in the General Specifications
must be applied.
The specifications also assume the 5080A is zeroed every seven days or whenever the ambient temperature changes by
more than 5 °C.
Warmup Time ......................................................... Twice the time since last warmed up, to a maximum of 30 minutes.
Settling Time ........................................................... Less than 7 seconds for all functions and ranges except as noted.
Standard Interfaces................................................. RS-232 and Ethernet
Temperature
Operating ............................................................ 0 °C to 50 °C
Calibration (tcal).................................................. 15 °C to 35 °C
Storage ............................................................... -20 °C to +70 °C
Temperature Coefficient ......................................... Temperature coefficient for temperatures outside tcal ±5 °C is 10 % of
the stated specification per °C for temperatures in the range of 0 °C to
35 °C. Above 35 °C, the temperature coefficient is 20 % of the stated
specification per °C.
Relative Humidity
Operating ............................................................ <80 % to 30 °C, <70 % to 40 °C, <40 % to 50 °C.
Storage ............................................................... <95 %, non-condensing
Altitude
Operating ............................................................ 2,000 m (6,500 ft) maximum
Non-operating ..................................................... 12,200 m (40,000 ft) maximum
Safety...................................................................... Meets EN 61010-1:2001, CAN/CSA-C22.2 No. 61010-1-04,
UL 61010-1:2004 Insulation Class I (bonded enclosure) Pollution
Degree 2 Indoor use only.
Analog Low Isolation............................................... 20 V
EMC ........................................................................ Meets EN 61326-1:2006.
Line Power
Line Voltage (selectable) .................................... 100 V, 120 V, 220 V, 240 V
Line Frequency ................................................... 47 to 63 Hz
Line Voltage Variation......................................... ±10 % about line voltage setting
Power Consumption................................................ 600 VA
Dimensions
Height.................................................................. 19.3 cm (7.6 in)
Width................................................................... 43.2 cm (17 in), 44.3 cm (17.5 in) including handles
Depth .................................................................. 53.8 cm (21.2 in)
Weight..................................................................... 22 kg (48 lb)
Specification Definition............................................ The specifications include stability, temperature coefficient, linearity,
line and load regulation, and the traceability of the external standards
used for calibration. It is not necessary to add anything to determine
the total specification for the temperature range indicated.
Specification Confidence Level............................... 99 %