5080A
Getting Started Manual
18
6. Replace the fuse compartment door and secure the door with the screws removed in
step 3.
Warming up the Calibrator
When you turn on the 5080A, allow a warm-up period of at least 30 minutes for the
internal components to stabilize. This ensures that the calibrator meets or exceeds the
specifications listed in this manual.
If you turn the 5080A Calibrator off after warm-up and then on again, allow a warm-up
period of at least twice the length of time it was turned off (maximum of 30 minutes). For
example, if the calibrator is turned off for 10 minutes and then on again, allow a warm-up
period of at least 20 minutes.
Front-Panel Features
Front-panel features (including all controls, displays, indicators, and terminals) are shown
in Figure 4. Each front-panel feature is described in Table 5.
Rear-Panel Features
Rear-panel features (including all terminals, sockets, and connectors) are shown in
Figure 5. Each rear-panel feature is described in Table 6.