Detailed Specifications
DC Voltage
Specification, tcal ±5 °C
±(% of output + μV)
Stability
Range
90 days 1 year
24 hours, ±1 °C
±(% of output + μV)
Resolution (μV)
Max Burden
[1]
0 to 329.999 mV 0.011 % + 10 0.013 % + 10 0.0035 % + 6 1
60 Ω
0 to 3.29999 V 0.008 % + 15 0.010 % + 15 0.0025 % + 10 10 300 mA
0 to 32.9999 V 0.008 % + 150 0.010 % + 150 0.0025 % + 100 100 600 mA
10 to 101.999 V 0.010 % + 1500 0.012 % + 1500 0.003 % + 1000 1000 300 mA
30 to 329.999 V 0.010 % + 1500 0.012 % + 1500 0.003 % + 1000 1000 120 mA
100 to 1020.00 V 0.010 % + 5500 0.012 % + 5500 0.003 % + 5000 10000 40 mA
Auxiliary Output (dual output mode only)
0 to 329.99 mV 0.10 % + 1000 0.12 % + 1000 0.03 % + 300 10 5 mA
0.33 to 3.2999 V 0.10 % + 1000 0.12 % + 1000 0.03 % + 300 100 5 mA
3.3 to 7.000 V 0.10 % + 1000 0.12 % + 1000 0.03 % + 300 1000 5 mA
[1] Remote sensing is not provided. Output resistance is 60 Ω for outputs <330 mV. Output resistance is <5 mΩ for outputs 0.33 V.
The AUX output has an output resistance of <1 Ω.
Noise
Range
Bandwidth 0.1 Hz to 10 Hz, p-p
±(ppm of output + floor)
Bandwidth 10 Hz to 10 kHz, rms
±( floor)
0 to 329.999 mV
0 + 3 μV 20 μV
0 to 3.29999 V
0 + 30 μV 200 μV
0 to 32.9999 V
0 + 300 μV
2 mV
10 to 101.999 V 30 + 5 mV 60 mV
30 to 329.999 V 30 + 5 mV 60 mV
100 to 1020.00 V 30 + 20 mV 100 mV
Auxiliary Output (dual output mode only)
0 to 329.99 mV
0 + 20 μV 60 μV
0.33 to 3.2999 V
0 + 200 μV 600 μV
3.3 to 7.000 V 0 + 2 mV 3 mV
DC Current
Specification, tcal ±5 °C
±(% of output + μA)
Range
90 days 1 year
Resolution
Max. Compliance
Voltage (V)
Max. Inductive
Load
0 to 329.99 μA
0.07 % + 0.1 0.075 % + 0.1 10 nA 9
0 to 3.2999 mA 0.06 % + 0.25 0.065 % + 0.25
0.1 μA
9
0 to 32.999 mA 0.048 % + 1.25 0.05 % + 1.25
1 μA
50
0 to 329.99 mA 0.048 % + 16.5 0.05 % + 16.5
10 μA
35
0 to1.0999 A
(in 3 A range)
0.14 % + 220 0.15 % + 220
100 μA
6
1.1 to 2.9999 A 0.18 % + 220 0.19 % + 220
100 μA
6
0 to 10.999 A
(in 20 A range)
0.23 % + 2500 0.25 % + 2500 1 mA 4
11 to 20.500 A
[1]
0.48 % + 3750 0.5 % + 3750 1 mA 4
2.5 H
[1] Duty Cycle: Currents <11 A may be provided continuously. For currents >11 A, the current may be provided 60-T-I minutes in any
60 minute period where T is the temperature in °C (room temperature is about 23 °C) and I is the output current in Amps. For
example, 17 A at 23 °C could be provided for 60-17-23 = 20 minutes each hour. When the 5080A is outputting currents between 5
and 11 amps for long periods, the internal self-heating reduces the duty cycle. Under those conditions, the allowable “on” time
indicated by the formula is achieved only after the 5080A is outputting currents <5 A for the “off” period first.