5080A
Getting Started Manual
16
Table 3. Line Power Cord Types Available from Fluke
Type Voltage/Current Fluke Option Number
North America
North America
Universal Euro
United Kingdom
Switzerland
Australia
South Africa
120 V/15 A
240 V/15 A
220 V/16 A
240 V/13 A
220 V/10 A
240 V/10 A
240 V/5 A
LC-1
LC-2
LC-3
LC-4
LC-5
LC-6
LC-7
LC-1 LC-2 LC-3 LC-4
LC-5 LC-6 LC-7
nn008f.eps
Figure 2. Line Power Cord Types Available from Fluke
Rack Mount Considerations
Rack mount to Calibrator using a support shelf or drawer. The Calibrator chassis is not
designed for use with front-mounted rack ears or side-mounted rack slides.
How to Replace the Current Fuses
The two Calibrator current outputs are protected by fuses. If the Calibrator is unable to
source current, one or both of the current fuses may be blown.
XW Warning
To avoid electric shock, turn the Calibrator off, remove the
power cord, and wait two minutes to allow the power
assemblies to completely discharge before opening the current
fuse access door.
To replace the current output fuses:
1. Turn off the Calibrator, remove the power cord, and wait two minutes for the power
assemblies to fully discharge.
2. Turn the Calibrator over.
3. Remove the two screws that hold the fuse compartment cover in place and remove
the cover as shown in Figure 3.