EasyManuals Logo

Fluke 5320A Getting Started Manual

Fluke 5320A
38 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #24 background imageLoading...
Page #24 background image
5320A
Getting Started Manual
10
AC/DC Current Uncertainty
Range Resolution
Uncertainty
±(% of Reading + mA)
300 mA 0.1 mA 0.15 % + 0.15
3 A 1 mA 0.15 % + 1.5
30 A 10 mA 0.30 % + 15
Phantom Power
Range ................................................................. 0 to 33 kVA
Resolution ......................................................... 3 digits
Uncertainty ........................................................
()()
22
uncunc
IV +
where V
unc
is specified uncertainty of measured voltage
and I
unc
is specified uncertainty of measured current.
Hipot Leakage Current Measurement Mode
Range………………………………………………0 to 300 mA ac rms or dc
Resolution…………………………………………4 1/2 digits
Frequency range…………………………………DC, 20 Hz to 400 Hz
Time constant…………………………………….1.5 s
Readings/second………………………………...2
Hipot Leakage Current Mode Uncertainty
Range Resolution
Uncertainty +/- (% of reading + μA)
300 uA
0.01 μA
0.3 % + 0.21
3 mA
0.1 μA
0.2 % + 1.5
30 mA
1 μA
0.2 % + 15
300 mA
10 μA
0.2 % + 150
Hipot Timer Measurement Mode
Range………………………………………………0.1 to 999 s
Resolution…………………………………………1 ms
Uncertainty ........................................................ 0.02 % + 2 ms (dc)
0.02 % + 20 ms (ac)
10 kV Adapter (1000:1 voltage divider)
Range ................................................................. 0 to 10 kV ac peak/dc
Resolution ......................................................... 4½ digits
Uncertainty ........................................................ 0.3 % of value + 5 V dc
0.5 % of value + 5 V ac at 50 or 60 Hz
80K-40 High Voltage Probe
Range ................................................................. 0 to 40 kV ac peak/dc
Resolution ......................................................... 4½ digits
Uncertainty ........................................................ 0.5 % of value + 10 V dc
0.5 % of value + 10 V ac at 50 or 60 Hz

Other manuals for Fluke 5320A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 5320A and is the answer not in the manual?

Fluke 5320A Specifications

General IconGeneral
BrandFluke
Model5320A
CategoryTest Equipment
LanguageEnglish

Related product manuals