EasyManua.ls Logo

Fluke 540B - Page 23

Fluke 540B
76 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
540B
STEP
CONTROL POSITION OPERATION
a
RANGE SHUNT
Connect
ac
source
under
test
or,
if
using
the
A45
Transfer
Swi
tch,
set
POWER
ON
the
OUTPUT MODE
switch
to
AC.
Tap
the
GAL V s
witch
agai
n
st
the
SENSITMTY
LOW
sp
ring
tension
to
MOMENTARY
just
long enough to
see
if
th
e GALVA-
MODE
OFF
NOMETER
needle
will
sta
y on
scale.
Adjust
the
COARSE
and
MEDIUM
GALV
REFERENCE
ADJUST
controls
until
a
reading
is
obtained.
REFERENCE
COARSE
ADJUST MEDIUM
b
GALV
OPEN
Adjust
the
GALV
ZERO
control
for
zero
reading
on
the
GALVA-
SENSITTITITY
MED
NOMETER.
GALV
ZERO
c GALV
LOCK
Adj
ust
the
MEDIUM
and
FINE
controls
for
a
zero
reading
on
the
GAL-
REFERENCE
MEDIUM VANOMETER.
ADJUST
FINE
d
GALV
OPEN
Adjust
the
GALV
ZERO
control
for
zero
reading
on
the
GALVA-
SENSITIVITY HIGH NOMETER.
GALV
ZERO
e
GALV LOCK
Adjust
the
FINE
control
for
a
zero
reading
on
the
GALVANOMETER.
REFERENCE
ADJUST FINE
f
GALV
OPEN
Disco
nnect
the
ac
circuit
under
test
from
the
current
shunt
and
connect
SENSITIVITY LOW
the
de
current
source,
using
care
not
to
exceed
the
c
urrent
rating
of
the
s hunt.
If
using
the
A45
Transfer
Switch,
se
t
the
OUTPUT MODE
s
witch
to
DC.
g
GALV
Tap
the
GALV
switch
against
the
spring
tension
to
MOMENTARY
just
long
enough
to
see
if
the
GALVANO
METER
needle
will
sta
y on
scale
whil
e
slowl
y
increasing
the
de
source
cur
ren
t
until
zero
is
reached
on
the
GALVANOMETER. Do
not
adjust
the
REFERENCE
ADJUST
con
-
trols
in
this
mode.
h GALV
LOCK
Fur
ther
refine
the
adjustment
of
de
current
so
urc
e
to
maintain
a
zero
SENSITIVITY MED
reading
on
the
GALVANOMETER
in
the
ME
D
and
HIGH
sensiti
vity
HIGH
positions
.
i
GALV
OPEN
Disco
nnect
the
de
source
and
connect
the
ac
current
to
be
measured
to
the
shunt.
H
using
the
A45
Transfer
Switch,
set
the
OUTPUT MODE
switch
to
AC .
j
GALV LOCK
Readjust
the
REFERENCE
ADJUST FINE
control
for
ze
ro
on
the
GAL-
VANOMETER.
k
GALV
OPE
N
Disconnec
t
the
ac
current
under
test
and
reconn
ect
the
de
source
to
the
shunt.
If
using
the
A45
Transfer
Switch,
set
the
O
UTPUT
MODE
s wi
tch
to
DC.
1 GALV LOCK
Adjust
the
de
current
source
to
maintain
a
zero
r
eading
on
the
GAL-
VANOMETER.
Measure
and
record
the
de
current
.
m
Repeat
steps
i
thro
ugh 1
at
least
three
times
or
until
repeatibility
is
attained.
Measure
and
record
the
de
c
urrent.
n
Reverse
the
input
polarity
to
the
thermocouple
by
depressing
the
POLARITY
switch
.
Repeat
steps
i
through
1
at
least
three
times
or
until
repeatibilit
y
is
attained.
Measur-e
and
record
the
de
current.
0
The
rms
v
alue
of unknown
ac
curren
t
is
equal
to
the
average
of
the
two
de
currents
recorded
in
steps
m
and
n .
NOTE:
If
a
test
report
from
a
sta
ndards
laboratory
is
used
,
calculate
the
unknown
ac
current
from
the
following equation:
=
(1
ao
I
Iac
+ 100
1
dc
Wh
ere:
Idc
=
average
value
of
the
two
de
current
s
from
step
o.
Iac
-
Idc
ao
=
100 =
ac
to
de
percent
difference
from
Idc
the
test
report.
Figure 2-
9.
MEASURING
AN UNKNOWN
AC
CURRENT
2-11