5560A/5550A/5540A
Service Manual
22
Figure 11. Amps DC with A40B-5A
35. Set the product to Standby. Disconnect the shunt. Connect the test equipment as shown in
Figure 12.
36. With no current applied from the Product to the Fluke A40B-10A shunt, proceed with the meter zero
function. On the 8588A, push ZERO, and then ZERO FUNC. Allow th zero function to finish.
37. Start with step 43. Set the output from the DUT, wait 2 minutes for settling and measure with the
meter and record the 8588A measurement in the DMM Reading on Shunt Output [V] column.
Complete both steps for the present connection (43 and 44). Refer to the Product manual for how
to set/use range lock on bottom of range outputs.
Figure 12. Amps DC with A40B-10A
5560A 8588A
INPUT
A40B-5A
OUTPUT
5560A
8588A
INPUT
A40B-10A
OUTPUT