5560A/5550A/5540A
Service Manual
44
Figure 22. Amps AC with A40B-100mA
28. From step 43, Table 16, set the output from the DUT/Measure with the meter and record the 8588A
measurement in the DMM Reading on Shunt Output [V] column.
29. Complete steps 43 and 44.
30. Disconnect the shunt output cable from the 8588A and connect to the 5790B Input 2.
31. Set the 5790B for Input 2, Auto range, High resolution ON, GUARD ON.
32. Connect a heavy gauge guard cable between the DUT and the 5790B.
33. From step 45, Table 16, output from the DUT/Measure with the 5790B and record the 5790B
measurement in the DMM Reading on Shunt Output [V] column. Complete steps 45 to 54.
34. Set the product to STBY.
35. Connect the test equipment as shown in Figure 23.
A40B-100mA
OUTPUT
5560A
8588A