Multifunction Calibrator
Full Verification
73
Table 15. 5790A Adjustment Counts
Product Range Nominal Value Measured Value Limits
2.2 mV .600 000 mV @ 1 kHz ±5.0 μV
2.2 mV 2.000 000 mV @ 10 Hz ±5.6 μV
2.2 mV 2.000 000 mV @ 20 Hz ±5.2 μV
2.2 mV 2.000 000 mV @ 40 Hz ±5.2 μV
2.2 mV 2.000 000 mV @ 1 kHz ±5.2 μV
2.2 mV 2.000 000 mV @ 20 kHz ±5.2 μV
2.2 mV 2.000 000 mV @ 50 kHz ±5.5 μV
2.2 mV 2.000 000 mV@100 kHz ±7.1 μV
2.2 mV 2.000 000 mV@300 kHz ±14.4 μV
2.2 mV 2.000 000 mV@500 kHz ±28 μV
2.2 mV 2.000 000 mV@1 MHz ±31.2 μV
22 mV 20.00 000 mV@10 Hz ±10.4 μV
22 mV 20.000 00 mV@20 Hz ±7.1 μV
22 mV 20.000 00 mV@40 Hz +/- 6.8 μV
22 mV 20.000 00 mV@1 kHz +/- 6.8 μV
22 mV 20.000 00 mV@20 kHz ±6.8 μV
22 mV 20.000 00 mV@50 kHz ±9.6 μV
22 mV 20.000 00 mV@100 kHz ±16.8 μV
22 mV 20.000 00 mV,300 kHz ±36.0 μV
22 mV 20.000 00 mV,500 kHz ±55.0 μV
22 mV 20.000 00 mV,1 MHz ± 87.0 μV
220 mV 200.000 0 mV,10 Hz ±69.0 μV
220 mV 200.000 0 mV, 20 Hz ±29.0 μV
220 mV 200.000 0 mV, 40 Hz ±21.2 μV
220 mV 200.000 0 mV, 1 kHz ±21.2 μV
220 mV 200.000 0 mV, 20 kHz ± 21.2 μV
220 mV 200.000 0 mV, 50 kHz ±36.0 μV
220 mV 200.000 0 mV, 100 kHz ±96.0 μV
220 mV 200.000 0 mV, 300 kHz ±165 μV
220 mV 200.000 0 mV, 500 kHz ±330 μV
220 mV 200.000 0 mV, 1 MHz ±640 μV