5730A
Calibration Manual
74
Table 15. 5790A Adjustment Counts (cont.)
Product Range Nominal Value Measured Value Limits
2.2 V .500 000 V, 40 Hz ±33 μV
2.2 V .500 000 V, 1 kHz ±33 μV
2.2 V .500 000 V, 20 kHz ±33 μV
2.2 V .500 000 V, 100 kHz ±88.5 μV
2.2 V .500 000 V, 300 kHz ±285 μV
2.2 V .500 000 V , 1 MHz ±1.3 mV
2.2 V 1.000 000 V, 40 Hz ±56 μV
2.2 V 1.000 000 V, 1 kHz ±56 μV
2.2 V 1.000 000 V, 20 kHz ±56 μV
2.2 V 1.000 000 V, 100 kHz ±137 μV
2.2 V 1.000 000 V, 300 kHz ±470 μV
2.2 V 1.000 000 V, 1 MHz ±2.2 mV
2.2 V 2.000 000 V, 10 Hz ±590 μV
2.2 V 2.000 000 V, 20 Hz ±220 μV
2.2 V 2.000 000 V, 40 Hz ±102 μV
2.2 V 2.000 000 V, 1 kHz ±102 μV
2.2 V 2.000 000 V, 20 kHz ±102 μV
2.2 V 2.000 000 V, 50 kHz ±166 μV
2.2 V 2.000 000 V, 100 kHz ±234 μV
2.2 V 2.000 000 V, 300 kHz ±840 μV
2.2 V 2.000 000 V, 500 kHz ±2.5 mV
2.2 V 2.000 000 V, 1 MHz ±4.0 mV
22 V 20.000 00 V, 10 Hz ±5.9 mV
22 V 20.000 00 V, 20 Hz ±2.2 mV
22 V 20.000 00 V, 40 Hz ±990 μV
22 V 20.000 00 V, 1 kHz ±990 μV
22 V 20.000 00 V, 20 kHz ±990 μV
22 V 20.000 00 V, 50 kHz ±1.66 mV
22 V 20.000 00 V, 100 kHz ±2.19 mV
22 V 20.000 00 V, 300 kHz ±6.6 mV