Appendices
Calibration Constant Information C
C-5
Group WBCABLE
Name Default Function
WBCABLE_CORR 1.0
The one and only correction for the cable, at
50 MHz
Group DC_2_2MV
Name Default Function
DI_2_2MV 5000.0 2_2MV range DAC volts per input volt
OF_2_2MV 0.0 2_2MV range DC offset
Group ZC_2_2MV
Name Default Function
Z_2_2MV 0.0 2_2MV range zero
SHO_2_2MV 0.0 2_2MV AUX input offset
IA_2_2MV 0.001 2_2MV range A/D volts per input volt
Group AC_2_2MV
Name Default Function
F1_2_2MV 1.0 2_2MV range flatness correction 1 (10 Hz)
F2_2_2MV 1.0 2_2MV range flatness correction 2 (100 Hz)
F3_2_2MV 1.0 2_2MV range flatness correction 3 (1 kHz)
F4_2_2MV 1.0 2_2MV range flatness correction 4 (10 kHz)
F5_2_2MV 1.0 2_2MV range flatness correction 5 (20 kHz)
F6_2_2MV 1.0 2_2MV range flatness correction 6 (50 kHz)
F7_2_2MV 1.0 2_2MV range flatness correction 7 (100 kHz)
F8_2_2MV 1.0 2_2MV range flatness correction 8 (300 kHz)
F9_2_2MV 1.0 2_2MV range flatness correction 9 (500 kHz)
F10_2_2MV 1.0 2_2MV range flatness correction 10 (800 kHz)
F11_2_2MV 1.0 2_2MV range flatness correction 11 (1 MHz)
Group DC_7MV
Name Default Function
DI_7MV 1000.0 7MV range DAC volts per input volt
OF_7MV 0.0 7MV range DC offset