5790B
Service Manual
C-6
Group ZC_7MV
Name Default Function
Z_7MV 0.0 7MV range zero
SHO_7MV 0.0 7MV AUX input offset
IA_7MV 0.00316228 7MV range A/D volts per input volt
Group AC_7MV
Name Default Function
F1_7MV 1.0 7MV range flatness correction 1 (10 Hz)
F2_7MV 1.0 7MV range flatness correction 2 (100 Hz)
F3_7MV 1.0 7MV range flatness correction 3 (1 kHz)
F4_7MV 1.0 7MV range flatness correction 4 (10 kHz)
F5_7MV 1.0 7MV range flatness correction 5 (20 kHz)
F6_7MV 1.0 7MV range flatness correction 6 (50 kHz)
F7_7MV 1.0 7MV range flatness correction 7 (100 kHz)
F8_7MV 1.0 7MV range flatness correction 8 (300 kHz)
F9_7MV 1.0 7MV range flatness correction 9 (500 kHz)
F10_7MV 1.0 7MV range flatness correction 10 (800 kHz)
F11_7MV 1.0 7MV range flatness correction 11 (1 MHz)
Group DC_22MV
Name Default Function
DI_22MV 500.0 22MV range DAC volts per input volt
OF_22MV 0.0 22MV range DC offset
Group ZC_22MV
Name Default Function
Z_22MV 0.0 22MV range zero
SHO_22MV 0.0 22MV AUX input offset
IA_22MV 0.01 22MV range A/D volts per input volt