Appendices
Calibration Constant Information C
C-19
Group WDC_220MV
Name Default Function
DI_220MV_WB 31.6228 220MV range, wideband input, DAC volts per input volt
IA_220MV_WB 3.16228 220MV range A/D volts per input volt
Group WAC_220MV
Name Default Function
F1_220MV_WB 1.0
220MV range, wideband input, flatness correction 1
(10 Hz)
F2_220MV_WB 1.0
220MV range, wideband input, flatness correction 2
(50 Hz)
F3_220MV_WB 1.0
220MV range, wideband input, flatness correction 3
(400 Hz)
F4_220MV_WB 1.0
220MV range, wideband input, flatness correction 4
(1 kHz)
F5_220MV_WB 1.0
220MV range, wideband input, flatness correction 5
(4 kHz)
F6_220MV_WB 1.0
220MV range, wideband input, flatness correction 6
(30 kHz)
F7_220MV_WB 1.0
220MV range, wideband input, flatness correction 7
(100 kHz)
F8_220MV_WB 1.0
220MV range, wideband input, flatness correction 8
(300 kHz)
F9_220MV_WB 1.0
220MV range, wideband input, flatness correction 9
(1 MHz)
F10_220MV_WB 1.0
220MV range, wideband input, flatness correction 10
(4 MHz)
F11_220MV_WB 1.0
220MV range, wideband input, flatness correction 11
(10 MHz)
F12_220MV_WB 1.0
220MV range, wideband input, flatness correction 12
(20 MHz)
F13_220MV_WB 1.0
220MV range, wideband input, flatness correction 13
(30 MHz)
F14_220MV_WB 1.0
220MV range, wideband input, flatness correction 14
(40 MHz)
F15_220MV_WB 1.0
220MV range, wideband input, flatness correction 15
(50 MHz)