5790B
Service Manual
3-16
Step Name Voltage to Apply Purpose of Calibration Step
LF (10 Hz) Linearity
LF (10 Hz) Linearity
2 V RMS, 10 Hz
600 mV RMS, 10 Hz
Generates a correction for
thermal sensor non-linearity at
low F and f
2.2 mV AC 10 Hz
2.2 mV AC 100 Hz
2.2 mV AC 1 kHz
2.2 mV AC 10 kHz
2.2 mV AC 20 kHz
2.2 mV AC 50 kHz
2.2 mV AC 100 kHz
2.2 mV AC 300 kHz
2.2 mV AC 500 kHz
2.2 mV AC 800 kHz
2.2 mV AC 1 MHz
2 mV RMS, 10 Hz
2 mV RMS, 100 Hz
2 mV RMS, 1 kHz
2 mV RMS, 10 kHz
2 mV RMS, 20 kHz
2 mV RMS, 50 kHz
2 mV RMS, 100 kHz
2 mV RMS, 300 kHz
2 mV RMS, 500 kHz
2 mV RMS, 800 kHz
2 mV RMS, 1 MHz
Generates flatness calibration
data for the 2.2 mV range
7 mV AC 10 Hz
7 mV AC 100 Hz
7 mV AC 1 kHz
7 mV AC 10 kHz
7 mV AC 20 kHz
7 mV AC 50 kHz
7 mV AC 100 kHz
7 mV AC 300 kHz
7 mV AC 500 kHz
7 mV AC 800 kHz
7 mV AC 1 MHz
6 mV RMS, 10 Hz
6 mV RMS, 100 Hz
6 mV RMS, 1 kHz
6 mV RMS, 10 kHz
6 mV RMS, 20 kHz
6 mV RMS, 50 kHz
6 mV RMS, 100 kHz
6 mV RMS, 300 kHz
6 mV RMS, 500 kHz
6 mV RMS, 800 kHz
6 mV RMS, 1 MHz
Generates flatness calibration
data for the 7 mV range
22 mV AC 10 Hz
22 mV AC 100 Hz
22 mV AC 1 kHz
22 mV AC 10 kHz
22 mV AC 20 kHz
22 mV AC 50 kHz
22 mV AC 100 kHz
22 mV AC 300 kHz
22 mV AC 500 kHz
22 mV AC 1 MHz
20 mV RMS, 10 Hz
20 mV RMS, 100 Hz
20 mV RMS, 1 kHz
20 mV RMS, 10 kHz
20 mV RMS, 20 kHz
20 mV RMS, 50 kHz
20 mV RMS, 100 kHz
20 mV RMS, 300 kHz
20 mV RMS, 500 kHz
20 mV RMS, 1 MHz
Generates flatness calibration
data for the 22 mV range
70 mV AC 10 Hz
70 mV AC 100 Hz
70 mV AC 1 kHz
70 mV AC 10 kHz
70 mV AC 20 kHz
70 mV AC 50 kHz
70 mV AC 100 kHz
70 mV AC 300 kHz
70 mV AC 500 kHz
70 mV AC 1 MHz
60 mV RMS, 10 Hz
60 mV RMS, 100 Hz
60 mV RMS, 1 kHz
60 mV RMS, 10 kHz
60 mV RMS, 20 kHz
60 mV RMS, 50 kHz
60 mV RMS, 100 kHz
60 mV RMS, 300 kHz
60 mV RMS, 500 kHz
60 mV RMS, 1 MHz
Generates flatness calibration
data for the 70 mV range
220 mV AC 10 Hz
220 mV AC 100 Hz
220 mV AC 1 kHz
200 mV RMS, 10 Hz
200 mV RMS, 100 Hz
200 mV RMS, 1 kHz
Generates flatness calibration
data for the 220 mV range