Calibration Adjustment and Verification
Calibration 3
3-17
Step Name Voltage to Apply Purpose of Calibration Step
220 mV AC 10 kHz
220 mV AC 20 kHz
220 mV AC 50 kHz
220 mV AC 100 kHz
220 mV AC 300 kHz
220 mV AC 500 kHz
220 mV AC 1 MHz
200 mV RMS, 10 kHz
200 mV RMS, 20 kHz
200 mV RMS, 50 kHz
200 mV RMS, 100 kHz
200 mV RMS, 300 kHz
200 mV RMS, 500 kHz
200 mV RMS, 1 MHz
700 mV AC 10 Hz
700 mV AC 100 Hz
700 mV AC 1 kHz
700 mV AC 10 kHz
700 mV AC 20 kHz
700 mV AC 50 kHz
700 mV AC 100 kHz
700 mV AC 300 kHz
700 mV AC 500 kHz
700 mV AC 1 MHz
600 mV RMS, 10 Hz
600 mV RMS, 100 Hz
600 mV RMS, 1 kHz
600 mV RMS, 10 kHz
600 mV RMS, 20 kHz
600 mV RMS, 50 kHz
600 mV RMS, 100 kHz
600 mV RMS, 300 kHz
600 mV RMS, 500 kHz
600 mV RMS, 1 MHz
Generates flatness calibration
data for the 700 mV range
2.2 V AC 10 Hz
2.2 V AC 100 Hz
2.2 V AC 1 kHz
2.2 V AC 10 kHz
2.2 V AC 20 kHz
2.2 V AC 50 kHz
2.2 V AC 100 kHz
2.2 V AC 300 kHz
2.2 V AC 500 kHz
2.2 V AC 1 MHz
2 V RMS, 10 Hz
2 V RMS, 100 Hz
2 V RMS, 1 kHz
2 V RMS, 10 kHz
2 V RMS, 20 kHz
2 V RMS, 50 kHz
2 V RMS, 100 kHz
2 V RMS, 300 kHz
2 V RMS, 500 kHz
2 V RMS, 1 MHz
Generates flatness calibration
data for the 2.2 V range
7 V AC 100 Hz
7 V AC 1 kHz
7 V AC 10 kHz
7 V AC 20 kHz
7 V AC 50 kHz
7 V AC 100 kHz
7 V AC 300 kHz
7 V AC 500 kHz
7 V AC 800 kHz
7 V AC 1 MHz
6 V RMS, 100 Hz
6 V RMS, 1 kHz
6 V RMS, 10 kHz
6 V RMS, 20 kHz
6 V RMS, 50 kHz
6 V RMS, 100 kHz
6 V RMS, 300 kHz
6 V RMS, 500 kHz
6 V RMS, 800 kHz
6 V RMS, 1 MHz
Generates flatness calibration
data for the 7 V range
22 V AC 100 Hz
22 V AC 1 kHz
22 V AC 10 kHz
22 V AC 20 kHz
22 V AC 50 kHz
22 V AC 100 kHz
22 V AC 300 kHz
22 V AC 500 kHz
22 V AC 1 MHz
20 V RMS, 100 Hz
20 V RMS, 1 kHz
20 V RMS, 10 kHz
20 V RMS, 20 kHz
20 V RMS, 50 kHz
20 V RMS, 100 kHz
20 V RMS, 300 kHz
20 V RMS, 500 kHz
20 V RMS, 1 MHz
Generates flatness calibration
data for the 22 V range
70 V AC 100 Hz
70 V AC 1 kHz
70 V AC 10 kHz
60 V RMS, 100 Hz
60 V RMS, 1 kHz
60 V RMS, 10 kHz
Generates flatness calibration
data for the 70 V range