Service Manual
3-54
Table 3-17. Wideband Verification Test Record
Step No.
5790B Range
Test Voltage
Frequency
1 Yr Flatness
Specification
Measured
Flatness
Error (%)
90 Days
Absolute
Error Spec
(%)
Measured
Absolute
Error (%)
1 7 V 3.2 V 10 Hz 0.10 0.31
2 7 V 3.2 V 20 Hz 0.10 0.31
3 7 V 3.2 V 50 Hz 0.03 0.31
4 7 V 3.2 V 105 Hz 0.03 0.31
5 7 V 3.2 V 200 Hz 0.03 0.31
7 7 V 3.2 V 2 kHz 0.03 0.31
8 7 V 3.2 V 10 kHz 0.03 0.31
9 7 V 3.2 V 20 kHz 0.03 0.31
10 7 V 3.2 V 50 kHz 0.03 0.31
11 7 V 3.2 V 120 kHz 0.03 0.31
12 7 V 3.2 V 200 kHz 0.03 0.31
13 7 V 3.2 V 500 kHz 0.03 0.31
14 7 V 3.2 V 700 kHz 0.05 No Spec No Spec
15 7 V 3.2 V 1 MHz 0.05 No Spec No Spec
16 7 V 3.2 V 1.2 MHz 0.05 No Spec No Spec
17 7 V 3.2 V 2 MHz 0.05 No Spec No Spec
18 7 V 3.2 V 3 MHz 0.10 No Spec No Spec
19 7 V 3.2 V 4 MHz 0.10 No Spec No Spec
20 7 V 3.2 V 6 MHz 0.10 No Spec No Spec
21 7 V 3.2 V 8 MHz 0.10 No Spec No Spec
22 7 V 3.2 V 9 MHz 0.10 No Spec No Spec
23 7 V 3.2 V 10 MHz 0.10 No Spec No Spec
24 7 V 3.2 V 12 MHz 0.15 No Spec No Spec
25 7 V 3.2 V 15 MHz 0.15 No Spec No Spec
26 7 V 3.2 V 17 MHz 0.15 No Spec No Spec
27 7 V 3.2 V 20 MHz 0.15 No Spec No Spec
28 7 V 3.2 V 23 MHz 0.35 No Spec No Spec
29 7 V 3.2 V 26 MHz 0.35 No Spec No Spec
30 7 V 3.2 V 28 MHz 0.35 No Spec No Spec
31 7 V 3.2 V 30 MHz 0.35 No Spec No Spec