Calibration Adjustment and Verification
Calibration Verification 3
3-55
Step No.
5790B Range
Test Voltage
Frequency
1 Yr Flatness
Specification
Measured
Flatness
Error (%)
90 Days
Absolute
Error Spec
(%)
Measured
Absolute
Error (%)
32 7 V 3.2 V 35 MHz 0.35 No Spec No Spec
33 7 V 3.2 V 40 MHz 0.35 No Spec No Spec
34 7 V 3.2 V 45 MHz 0.35 No Spec No Spec
35 7 V 3.2 V 50 MHz 0.35 No Spec No Spec
36 2.2 V 1.0 V 10 Hz 0.10 0.33
37 2.2 V 1.0 V 20 Hz 0.10 0.33
38 2.2 V 1.0 V 50 Hz 0.03 0.33
39 2.2 V 1.0 V 105 Hz 0.03 0.33
40 2.2 V 1.0 V 200 Hz 0.03 0.33
42 2.2 V 1.0 V 2 kHz 0.03 0.33
43 2.2 V 1.0 V 10 kHz 0.03 0.33
44 2.2 V 1.0 V 20 kHz 0.03 0.33
45 2.2 V 1.0 V 50 kHz 0.03 0.33
46 2.2 V 1.0 V 120 kHz 0.03 0.33
47 2.2 V 1.0 V 200 kHz 0.03 0.33
48 2.2 V 1.0 V 500 kHz 0.03 0.33
49 2.2 V 1.0 V 700 kHz 0.05 No Spec No Spec
50 2.2 V 1.0 V 1 MHz 0.05 No Spec No Spec
51 2.2 V 1.0 V 1.2 MHz 0.05 No Spec No Spec
52 2.2 V 1.0 V 2 MHz 0.05 No Spec No Spec
53 2.2 V 1.0 V 3 MHz 0.10 No Spec No Spec
54 2.2 V 1.0 V 4 MHz 0.10 No Spec No Spec
55 2.2 V 1.0 V 6 MHz 0.10 No Spec No Spec
56 2.2 V 1.0 V 8 MHz 0.10 No Spec No Spec
57 2.2 V 1.0 V 9 MHz 0.10 No Spec No Spec
58 2.2 V 1.0 V 10 MHz 0.10 No Spec No Spec
59 2.2 V 1.0 V 12 MHz 0.15 No Spec No Spec
60 2.2 V 1.0 V 15 MHz 0.15 No Spec No Spec
61 2.2 V 1.0 V 17 MHz 0.15 No Spec No Spec
62 2.2 V 1.0 V 20 MHz 0.15 No Spec No Spec
63 2.2 V 1.0 V 23 MHz 0.35 No Spec No Spec
64 2.2 V 1.0 V 26 MHz 0.35 No Spec No Spec