IR Thermometer
Specifications
29
Specifications
Temperature Range
-30
°
C to +600
°
C
Accuracy
(Calibration geometry with ambient
temperature 23 °C ±2 °C)
0
°
C:
±
1
°
C or
±
1 % of reading, whichever is
greater
-10 °C to <0 °C: ±2 °C
< -10
°
C:
±
3
°
C
Response Time (95 %)
<500 ms (95 % of reading)
Spectral Response 8 microns to 14 microns
Emissivity
0.10 to 1.00
Temperature Coefficient
±0.1 °C/°C or ±0.1 %/°C of reading (whichever is
greater)
Optical Resolution
20:1
(calculated at 90 % energy)
Display Resolution
0.1
°
C
Repeatability (% of reading)
±
0.5 % of reading or
±
0.5
°
C, whichever is greater
Power
1 AA IEC LR06 Battery
Battery Life 30 hours with laser and backlight on
1.888.610.7664 info@Fluke-Direct.com
Fluke-Direct.com