714C
Users Manual
28
Specifications
Specifications are based on a one year calibration cycle
and apply from +18 °C to +28 °C unless stated otherwise.
All specifications assume a 5 minute warmup period.
Thermocouple mV Input
Range Resolution
Accuracy (% of Reading + Floor)
1 year 2 year
-10 mV to 75 mV 0.001 mV 0.015 % + 10 µV 0.02% + 15 µV
Temperature coefficient: ±(0.002% of reading + 0.002 % of range) /°C ( <18°C or >28°C)
Thermocouple mV Output
Range Resolution Accuracy (% of Output +Floor)
1 year 2 year
-10 mV to 75 mV 0.01 mV 0.015 % + 10 µV 0.02 % + 15 µV
Temperature Coefficient: ± (0.002 % of output + 0.002 % of range) /°C ( <18 °C or >28 °C)
mA Input
Range Resolution
Accuracy (% of Reading + Floor)
1 year 2 year
0 mA to 24 mA 0.001 mA 0.01 % + 2 µA 0.02 % + 4 µA
Temperature Coefficient: ± (0.002 % of output + 0.002 % of range) /°C ( <18 °C or >28 °C)