Thermocouple Calibrator
Specifications
31
G
100 to 300 1.6 2.4 1.2 1.8
300 to 1500 1.0 1.5 1.0 1.5
1500 to 2315 2.0 3.0 1.6 2.4
D
0 to 300 1.6 2.4 1.2 1.8
300 to 1500 1.0 1.5 1.0 1.5
1500 to 2315 2.0 3.0 1.6 2.4
P
0 to 1000 1.6 2.4 0.6 0.9
1000 to 1395 2.0 3.0 0.8 1.2
M
-50 to 100 1.0 1.5 0.4 0.6
100 to 1000 1.6 2.4 0.6 0.9
1000 to 1410 2.0 3.0 0.8 1.2
1) Sensor inaccuracies not included.
2) Accuracy with external cold junction; for internal junction add 0.2 °C
3) Temperature scale: ITS-90
Compensation: NIST Monograph 175 for B, R, S, E, J, K, N, T. DIN 43710 for L, U. GOST P 8.585-2001 (Russia) for
BP and XK. ASTM E988-96 for C (W5Re/W26Re). ASTM E1751/E1751M 09ԑ1 for G, D, P, M
4) Resolution: 0.1 °C
5) Temperature Coefficient: 0.05 °C/°C ( <18 °C or >28 °C )
0.07 °C/°C for C type >1800 °C and for BP type >2000 °C
Loop Power Supply
Open circuit voltage: 24 V ±10 %
Loaded circuit voltage: 18 V minimum at 20 mA current
Short circuit protected