EasyManua.ls Logo

Fluke 725/FPC - Page 56

Fluke 725/FPC
68 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
725
Users Manual
48
Table 8B. Remote Control Lower Display (cont)
Serial Input Description
0-9 Enter a source value using ascii characters 0,1,2,...9,-,.terminated by <CR> (carriage return)
-,.
<CR>
b Single Broadcast of most recent lower display value and units
Table 8C. "S" Commands Select Sensor Type
Selection Entry
Serial
Input
No. Thermocouple Type RTD Type
S 1 J Pt100 (3926)
2 K Pt100 (385)
3 T Pt100 (3916)
4 E Pt200 (385)
5 R Pt500 (385)
6 S Pt1000 (385)
7 B Ni120
8L
9U
AN
BmV
1.800.561.8187information@ITM.com