Page 15
Product Disposal
Dispose of the Product in a professional and environmentally sound
manner:
1. Delete personal data on the Product before disposal.
2. Remove batteries that are not integrated into the electrical system
before disposal and dispose of batteries separately.
3. If this Product has an integral battery, put the entire Product in the
electrical waste.
Specifications
Electrical Specifications
DC Current Measurement
With Jaw
Ranges ............................................... 0 mA to 20.99 mA;
21 mA to 100 mA
Resolution........................................... 0.01 mA, 0.1 mA
Accuracy ............................................. 0.2 % + 5 counts, 1 % + 5 counts
In Circuit
Range ................................................. 0 mA to 24 mA
Resolution........................................... 0.01 mA
Accuracy ............................................. 0.2 % + 2 counts
Current Source
Range ................................................. 0 mA to 24 mA
Resolution........................................... 0.01 mA
Accuracy ............................................. 0.2 % + 2 counts
mA Drive ............................................. 24 mA into 1000 Ω
Current Simulate
Range ................................................. 0 mA to 24 mA
Resolution........................................... 0.01 mA
Accuracy ............................................. 0.2 % + 2 counts
Maximum Voltage ............................... 50 V
DC Voltage Measurement (773/773-II)
Range ................................................. 0 V to 30 V
Resolution........................................... 0.01 V
Accuracy ............................................. 0.2 % + 2 counts
DC voltage source (773/773-II)
Range ................................................. 0 V to 10 V
Resolution........................................... 0.01 V
Accuracy ............................................. 0.2 % + 2 counts
mA Drive ............................................. 2 mA max all conditions
mA IN/OUT (773/773-II)
Sourcing range ................................... 0 mA to 24 mA

Other manuals for Fluke 773

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 773 and is the answer not in the manual?

Fluke 773 Specifications

General IconGeneral
Operating altitude0 - 2000 m
Storage temperature (T-T)-25 - 60 °C
Operating temperature (T-T)-10 - 50 °C
Operating relative humidity (H-H)0 - 90 %
Battery typeAA
Power sourceBattery
Battery voltage1.5 V
Battery technologyAlkaline
Number of batteries supported4
Product typeDigital multimeter
Product colorGrey, Red, Yellow
DC current range0 - 0.024 A
DC voltage range0 - 30 V
Placement supportedHandheld multimeter
Basic accuracy (DC current)0.2% + 2 counts
Weight and Dimensions IconWeight and Dimensions
Depth44 mm
Width70 mm
Height246 mm
Weight410 g