80218
6-34. Specifications
SHUNT
10 amps at 100 mV
ACCURACY (
I8°C
to
28°C)
DC to
10 kHz
±0.25%
10 kHz-100 kHz Rising
to
1 dB at 100 kHz typical
TEMPERATURE COEFFICIENT
0.005%/°C
INDUCTANCE 8.3
nH in series w/0.01fl shunt
OVERLOAD Up to one
minute
at 20A
with
a
1/4
duty cycle for
recovery after currents between 10 A and 20A
CONNECTS
TO
3/4 inch
center banana
jacks
CONNECTORS
5-way
binding posts (red
and
black)
6-35.
AC/DC
CURRENT PROBE (Y8100)
6-36.
intro
duciion
6-37. The Fluke Y8100
AC/
DC Current Probe is a clamp-on probe
that
is used with a
voltmeter, multimeter,
or
oscilloscope to read dc, ac, or
composite (aeon
dc)
currents.
The
jaws on the Y8100 are designed
to clamp around conductors up to
3/4
inch in diameter.
The
pistol shape allows safe, easy,
one-hand operation when
making current
measurement. The Model
Y8
100
probe is battery
powered
with size AA cells. It measures
current
to 200A dc or ac rms using
almost any voltmeter. Two
ranges, 20A and 200A,
produce a
2V output at full-range current.
6-38. Specifications
RANGES 20A ac
or
dc
200A ac or dc
RATED OUTPUT
2V at full range
ACCURACY
DC to 200 Hz ±2%
of range
200
Hz to 1 kHz
<100A
add
±3%
reading
>100A add
±6%
reading
CALIBRATION
CYCLE
1
year
FREQUENCY
RESPONSE
dc to 1.0 kHz
RECOMMENDED LOAD >3.0
kfl
TEMPERATURE
RANGE
+ 15°C
to
+35°C; for specified accuracy
-10°C
to
+50°C; storage and
operation at reduced
accuracy.
HEATING
LIMITATION
Prolonged operation
above 200Aacor I kHz
can
cause
damage to the Y8100.
WORKING VOLTAGE
RATING .
Core to output; 600V
dc or 480V ac maximum
output
to ground; 42V dc or 30V ac
APERTURE SIZE 3/4"
(19
mm) diameter
SIZE-OVERALL
9"x
4-1/2"
x 1-7/
16"
(230
mm x i 15 mm x 37 mm)
WEIGHT 14 ounces (0.4 kg), with batteries
POWER
Battery, four AA cells
BATTERY
LIFE
Aikaline-20 hours
continuous
6-39.
AC CURRENT TRANSFORMER (Y8101)
6-40.
Introduction
6-41.
The Model Y810I (Figure 6-1
)
is a small
clamp-on current
transformer designed to
6-7