87V MAX 
Users Manual 
46 
Duty Cycle (Vdc and mVdc) 
Range Accuracy 
0.0 % to 99.9 %
 [1]
 
Within ± (0.2 % per kHz + 0.1 %) for rise times < 1 μs. 
[1]  0.5 Hz to 200 kHz, pulse width >2 μs. Pulse width range is determined by the frequency by the frequency of the signal. 
Input Characteristics 
Function 
Overload 
Protection
[1]
 
Input 
Impedance 
(nominal) 
Common Mode  
Rejection Ratio  
(1 kΩ unbalance) 
Normal Mode Rejection 
 
1000 V rms 
10 MΩ <100 pF 
> 120 dB at dc, 50 Hz 
or 60 Hz 
> 60 dB at 50 Hz or 60 Hz 
 
1000 V rms 
> 120 dB at dc, 50 Hz 
or 60 Hz 
> 60 dB at 50 Hz or 60 Hz 
 
1000 V rms 
10 MΩ < 100 pF 
(ac-coupled) 
> 60 dB, dc to 60 Hz 
 
 
 
Open Circuit 
Test Voltage 
Full Scale Voltage  Typical Short Circuit Current 
 
 
To 6 MΩ 
5 MΩ or 
60 nS 
600 Ω 6 kΩ 60 kΩ 600 kΩ 6 MΩ 50 MΩ 
Ω 
1000 V rms  <2.8 V dc  <850 mV dc  <1.3 V dc 
500 μA 100 μA 10 μA 1 μA 0.2 μA 0.1 μA 
 
1000 V rms  <2.8 V dc  2.200 V dc  1.0 mA typical 
[1] 10
6
 V Hz Max 
 
1.800.561.8187info@Fluke-Direct.ca
Fluke-Direct.ca