EasyManuals Logo

Fluke 8846A User Manual

Fluke 8846A
104 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #19 background imageLoading...
Page #19 background image
Introduction and Specifications
General Specifications 1
1-9
Model/Fluke PN Description
1258730 Agreement, Calibration, Z540 Traceable, without data
Table 1-5. Accessories (cont)
Model/Fluke PN Description
1259340 Agreement, Calibration, Accredited
2441827 Agreement, Calibration, primary standards lab
1540600 Agreement, Calibration, artifact
General Specifications
Power
Voltage
100 V Setting ...................................................... 90 V to 110 V
120 V Setting ...................................................... 108 V to 132 V
220 V Setting ...................................................... 198 V to 242 V
240 V Setting ...................................................... 216 V to 264 V
Frequency ............................................................... 47 Hz to 440 Hz. Automatically sensed at power-on.
Power Consumption................................................ 28 VA peak (12 Watt average)
Dimensions
Height...................................................................... 88 mm (3.46 in.)
Width....................................................................... 217 mm (8.56 in.)
Depth ...................................................................... 297 mm (11.7 in.)
Weight..................................................................... 3.6 kg (8.0 lb)
Shipping Weight...................................................... 5.0 kg (11.0 lb)
Display
Vacuum Fluorescent Display, dot matrix
Environment
Temperature
Operating ............................................................ 0 °C to 55 °C
Storage ............................................................... -40 °C to 70 °C
Warm Up............................................................. 1 hour to full uncertainty specifications
Relative Humidity (non-condensing)
Operating ............................................................ 0 °C to 28 °C <90 %
28 °C to 40 °C <80 %
40 °C to 55 °C <50 %
Storage ............................................................... -40 °C to 70 °C <95 %
Altitude
Operating ............................................................ 2,000 Meters
Storage ............................................................... 12,000 Meters
Vibration and Shock................................................ Complies with Mil-T-28800F Type III, Class 5 (Sine only)
Safety
Designed to comply with IEC 61010-1:2000-1, UL 61010-1A1, CAN/CSA-C22.2 No. 61010.1, CAT I 1000V/CAT II 600V

Table of Contents

Other manuals for Fluke 8846A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 8846A and is the answer not in the manual?

Fluke 8846A Specifications

General IconGeneral
Frequency Range3 Hz to 300 kHz
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Resistance Range10 Ω to 1 GΩ
ContinuityYes
Diode TestYes
Temperature MeasurementYes
Display5.5 digit
ConnectivityUSB, RS-232, GPIB
DC Current Range100 µA to 10 A
AC Current Range100 µA to 10 A
Frequency Accuracy0.01 %
Capacitance Range1 nF to 100 µF
Display TypeVFD
Power Supply100 V to 240 V
AC Voltage Accuracy±(0.06% + 0.03% of reading)
Resistance Accuracy±(0.01% + 0.001% of reading)

Related product manuals