8845A/8846A
Users Manual
4-14
2. With the leads open, press Z.
3. Connect the Meter’s test leads as shown in Figure 4-6.
2
INPUT SENSE
HI
LO
FRONT
REAR
10 A
1V
00
mA
4
W
HI
LO
2W/4W
V
300V
FUSED
1000 V CAT I
600V CAT II
4
caw027.eps
Figure 4-6. Measuring Capacitance
Refer to the “Range Keys” section in Chapter 3 of this manual for information on how to
adjust the measurement range.
Measuring RTD Temperature (8846A only)
The Fluke 8846A is capable of measuring temperatures between -200 °C and 600 °C
using Resistance Temperature Detectors (RTDs).
To make a temperature measurement:
1. Connect the RTD to the Input HI, and LO, then the SENSE HI and LO connectors
as shown in Figure 4-7.
2
INPUT SENSE
HI
LO
FRONT
REAR
10 A
1V
00
mA
4
W
HI
LO
2W/4W
V
300V
FUSED
1000 V CAT I
600V CAT II
4
caw028.eps
Figure 4-7. Temperature Measurements
2. Press T to display the measured temperature as shown below.