8845A/8846A
Users Manual
4-18
Setting Number of Samples per Trigger
When in external trigger mode, the Meter will take between 1 and 50,000 measurements
for each trigger received.
To set the number of samples, or measurements the Meter will take for each external
trigger received:
1. Press Lto display the measurement setup menu.
2. Press the TRIGGER soft key.
3. Press the soft key labeled #SAMPLES.
4. Use the soft keys to set the number of samples between 1 and 50,000.
Select the digit in the value by pressing either <-- or -->.
With the digit selected, press the soft key labeled -- to decrement the digit or ++ to
increment the character.
5. Press ENTER.
Connecting an External Trigger
The TRIG I/O jack on the Meter’s rear panel is used to connect to an external trigger
signal. The falling edge of a TTL signal will trigger the Meter to start taking
measurements, if the Meter is in external trigger mode.
Figure 4-9 indicates the purpose of the three pins in the TRIG I/O connector.
TRIG I/O
Pin 3: Measure Complete OutPin 1: Trigger In
Pin 2: Signal Ground
caw059.eps
Figure 4-9. TRIG I/O Pin-out Description
Monitoring the Measurement-Complete Signal
In addition to being a trigger input, the TRIG I/O jack on the Meter’s rear panel also
provides a signal indicating the completion of a measurement cycle. A falling edge of a
TTL signal indicates a measurement cycle is complete. See Figure 4-9 above to identify
which pins on the TRIG I/O connector are used to sense the measurement-complete
signal.