8845A/8846A
Users Manual
4-6
Function modifiers:
Filter Displays the filter menu. For best accuracy and stable readings,
choose a filter based on the lowest frequency to be measured and the
performance needed.
3HZ SLOW Provides higher measurement accuracy on ac signals
between 3 Hz and 20 Hz. However, the measurement
cycle time is longer than it is when using the 20 Hz
filter.
20HZ Provides higher measurement accuracy on ac signals
between 20 Hz and 200 Hz. The measurement cycle
time is longer, however, than it is when using the
200 Hz filter.
200HZ Provides accurate measurements on ac signals 200 Hz
and above.
dB Displays the measured voltage as a decibel value referenced to a
stored relative value (dB = 20 log(Vnew/Vstored). The stored value is
obtained from the first measurement the Meter makes after pressing
the dB soft key. All future measurements are displayed using the
stored value as an offset. To take the Meter out of dB mode, press the
dB soft key.
dB Displays the measured voltage as a power decibel value referenced to
1 milliwatt (dBm = 10 log(power of Vnew into the reference
resistance / 1 mW) or 10 log(V
2
/R*1 mW) where R is the resistance.
To accommodate the various impedances that a dBm measurement
can be made across, the Meter allows for the selection of 21 different
impedance values.
To set the dB reference impedance:
1. Press L.
2. Press the soft key labeled dB Ref.
The available impedance settings are presented in sets of three
values. To move to a higher set of impedance values, press ++
-->. Press <-- -- to go to a lower set of impedance values.
3. With the impedance highlighted, press the soft key under the
value selected.
2ND MEAS Cycles the secondary display through the measurement functions
listed below, and then off. When a second measurement function is
selected, the 2ND MEAS soft key label is highlighted.
VDC - Displays the dc voltage on which the ac signal may be riding.
Note
ACV/DCV dual mode should not be used at frequencies
below 10 Hz.
Frequency - Displays the frequency of the ac signal applied to the
Input HI and LO connectors of the Meter.

Table of Contents

Other manuals for Fluke 8846A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 8846A and is the answer not in the manual?

Fluke 8846A Specifications

General IconGeneral
Frequency Range3 Hz to 300 kHz
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Resistance Range10 Ω to 1 GΩ
ContinuityYes
Diode TestYes
Temperature MeasurementYes
Display5.5 digit
ConnectivityUSB, RS-232, GPIB
DC Current Range100 µA to 10 A
AC Current Range100 µA to 10 A
Frequency Accuracy0.01 %
Capacitance Range1 nF to 100 µF
Display TypeVFD
Power Supply100 V to 240 V
AC Voltage Accuracy±(0.06% + 0.03% of reading)
Resistance Accuracy±(0.01% + 0.001% of reading)

Related product manuals