Fluke 8846A - Standard Event Status and Standard Event Status Enable Registers

Fluke 8846A
102 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Digital Multimeter
Service Requests (IEEE 488 Only) and Status Registers
23
Voltage Overload
2
0
= 1 2
8
= 256
2
1
= 2 2
9
= 512
2
2
= 4 2
10
= 1024
2
3
= 8 2
11
= 2048
2
4
= 16 2
12
= 4096
2
5
= 32 2
13
= 8192
2
6
= 64 2
14
= 16384
2
7
= 128 2
15
= 32768
Questionable Data Binary Wieghts
Event Register Enable Register
STAT:QUES:EVEN? STAT:QUES:ENAB <value>
STAT:QUES:ENAB?
"OR"
0
Current Overload1
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
Ohms Overload
Not Used
9
Limit Test Fail LO11
Limit Test Fail HI12
13
Remote Mode
15
Not Used
Not Used
Not Used
Operation Complete
Standard Event
Event Register Enable Register
*ESR? *ESE <value>
*ESE?
"OR"
0
Not Used
Query Error2
Device Error3
Execution Error4
5
Command Error
7
Not Used
Power On
Not Used
Status Byte
Output Buffer
Summary Register Enable Register
Serial Poll (SPOLL)
*STB?
*SRE <value>
*SRE?
"OR"
0
Not Used
Not Used
Questionable Data
3
Message Available
4
5
Standard Event
7
6
Request Service
Not Used
caw01f.eps
Figure 3. Overview of Status Data Structure
Standard Event Status and Standard Event Status Enable Registers
The Standard Event Status Register (ESR) assigns events to specific bits (see Figure 3
and Table 5). When a bit in the ESR is set (for example, 1), the event that corresponds to
that bit has occurred since the register was last read or cleared. For example, if bit 3
(DDE) is set to 1, a device-dependent error has occurred.
The Standard Event Status Enable Register (ESE) is a mask register that allows the host
to enable or disable (mask) each bit in the ESR. When a bit in the ESE is 1, the
corresponding bit in the ESR is enabled. When any enabled bit in the ESR changes from
O to 1, the Standard Event bit in the Status Byte Register also goes to 1. When the ESR is
read (using the *ESR? command) or cleared (using the *CLS command), the Standard
Event bit in the Status Byte Register returns to 0.