EasyManua.ls Logo

Fluke 8846A - Programming the Trigger System

Fluke 8846A
102 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8845A/8846A
Programmers Manual
54
Parameters <n> = “CALCulate” Enable the storing of measurements.
“” Disable the storage of measurements.
Example DATA:FEED RDG_STORE, “CALC”.
DATA:FEED?
Description Retrieves the state of the reading store setting.
Query DATA:FEED? Returns CALC or “”.
CALCulate:KMATh:MMFactor(?) <value>
Description Sets the value in the “m” register for the mx+b calculation.
Parameters <value> = 0 to ±999.999999
Examples CALC:KMAT:MMF 37 Sets “m” register to 37.
Query CALC: KMAT:MMF? Retrieves the “m” register value.
CALCulate:KMATh:MBFactor(?) <value>
Description Sets the value in the “b” register for the mx+b calculation.
Parameters <value> = 0 to ±999.999999
Examples CALC:KMAT:MBF 37 Sets “b” register to 37.
Query CALC: KMAT:MBF? Retrieves the “b” register value.
CALCulate:KMATh:MUNits(?) <name> = 3 characters using ‘A’ through ‘Z’
Description Sets the units for the mx+b calculation.
Parameters <name> = A to ZZZ Up to three letters
Examples CALC:KMAT:MUN VOL Sets units to “VOL”.
Query CALC: KMAT:MUN? Retrieves the set units value.
CALCulate:KMATh:STATe(?) <n>
Description Enables or disables the mx+b calculation.
Parameters <n> = ON or 1 Enables mx+b calculations.
OFF or 0 Disable mx+b calculations
Examples CALC:KMAT:STAT ON Enables mx+b calculations.
Query CALC: KMAT:STAT? Retrieves the state of the mx+b calculation (0 =
OFF, 1 = ON).
Programming the Trigger System
The Meter is designed to provide extensive control over the triggering of a single
measurement or series of measurements. One trigger can cause the Meter to take a single
measurement or multiple measurements up to 50,000 per trigger.
Operating the Meter remotely requires selecting one of three trigger methods: internal,
external, or bus triggering. Once a source is selected, the Meter must be in a “wait-for-
trigger” state before the trigger will start a measurement cycle.
Figure 4 shows the multimeter’s triggering system.

Table of Contents

Other manuals for Fluke 8846A

Related product manuals