8845A/8846A
Programmers Manual
68
Table 20. IEEE Common Commands (cont.)
Command Name Description
*IDN? Identification Query Meter returns the identification code of the meter as four fields
separated by commas. These fields are: manufacturer
("FLUKE"); model (“45"); seven-digit serial number; version of
main software and version of display software. (example:
FLUKE, 45, 9080025, 2.0, D2.0)
*OPC Operation Complete
Command
Meter sets the Operation Complete bit in the Standard Event
Status Register when parsed.
*OPC? Operation Complete
Query
Meter places an ASCII "1 " in the output queue when parsed.
*RST Reset Meter performs power-up reset, except that the state of IEEE
488 interface is unchanged, including: instrument address,
Status Byte, and Event Status Register.
*SRE <value>
Service Request
Enable
Sets the "Service Request Enable Register" to <value>, an
integer between 0 and 255. The value of bit six is ignored
because it is not used by the Service Request Enable
Register.
Note that <value> is an integer whose binary equivalent
corresponds to the state (1 or 0) of bits in the register. If
<value> is not between 0 and 255, an Execution Error is
generated.
*SRE?
Service Request
Enable Query
Meter returns the <value> of the Service Request Enable
Register (with bit six set to zero); <value> is an integer whose
binary equivalent corresponds to the state (1 or 0) of bits in the
register.
*STB? Read Status Byte Meter returns the <value> of the Status Byte with bit six as the
Master Summary bit.
<value> is an integer whose binary equivalent corresponds to
the state (1 or 0) of bits in the register.
*TRG Trigger Causes the meter to trigger a measurement when parsed.
*WAI Wait-to-continue Command required by IEEE 488.2 standard. Non-operational
in Fluke 45 Dual Display Multimeter. Command accepted but
has no effect.

Table of Contents

Other manuals for Fluke 8846A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 8846A and is the answer not in the manual?

Fluke 8846A Specifications

General IconGeneral
Frequency Range3 Hz to 300 kHz
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Resistance Range10 Ω to 1 GΩ
ContinuityYes
Diode TestYes
Temperature MeasurementYes
Display5.5 digit
ConnectivityUSB, RS-232, GPIB
DC Current Range100 µA to 10 A
AC Current Range100 µA to 10 A
Frequency Accuracy0.01 %
Capacitance Range1 nF to 100 µF
Display TypeVFD
Power Supply100 V to 240 V
AC Voltage Accuracy±(0.06% + 0.03% of reading)
Resistance Accuracy±(0.01% + 0.001% of reading)

Related product manuals