COMMAND REFERENCE 4 - 19
*
IDN? Identification
Syntax:
*
IDN?
Response:
<manufacturer>,<model>,<serial_number>,<sw_level>
<manufacturer> E.g., FLUKE
<model> E.g., PM3394B
<serial_number> Always 0
<sw_level> <sw_id>:<mask_id>:<UFO_id>
<sw_id> Firmware identification, consisting of:
- Software type, e.g., SW3394BIM
(I=IEEE, M=Math Plus)
- Software version, e.g., V4.0
- Software date (year-month-day)
<mask_id> Mask identification, e.g., UHM V1.0
<UFO_id> UFO identification, e.g., UFO V2.0
Description:
The
*
IDN? query reports the identification of the instrument. The response to the
*
IDN? query consists of the fields above in Arbitrary ASCII Response Data
format. This implies that the
*
IDN? query must be the last query in a program
message unit, because the arbitrary ASCII response data is terminated with the
New Line character (10 decimal).
The <sw_id> parameter identifies the type, version, and date of the instrument
firmware.
The <mask_id> parameter identifies the version of the Universal Host Mask
processor software.
The <UFO_id> parameter identifies the version of the Universal Front processor
software.
Example:
Send
*
IDN?
Read
FLUKE,PM3384B,0,SW3394BIM V4.0 1996-10-02:UHM V1.0:UFO V2.0
Front panel compliance:
The
*
IDN? query is the remote equivalent of the Maintenance option of the
UTILITY menu.