COMMAND REFERENCE 4 - 25
*
SAV Save instrument setup
Syntax:
*
SAV <numeric_data>
Description:
The
*
SAV command saves the current instrument settings into one of the internal
memory registers 1 .. 10. The settings in memory register 0 are standard settings,
which can only be recalled. The settings in the memory registers 0 through 10 can
be recalled by sending the
*
RCL command.
Example:
Send
*
SAV 2 Stores the actual instrument settings into
memory register 2.
.
.
Send
*
RCL 2 Restores the instrument settings from memory
register 2.
Front panel compliance:
The
*
SAV/
*
RCL commands are the remote equivalent of the front panel softkey
operation via the SETUPS/RECALL menu. The standard settings stored in
memory 0 can be changed via the front panel FRONT SETUPS menu.