4 - 28 COMMAND REFERENCE
*
TRG Trigger
Syntax:
*
TRG
Description:
The
*
TRG command triggers the instrument by generating a Group Execute
Trigger (
GET
) code.
Example:
Send *RST Resets the instrument.
Send
TRIGger:SOURce BUS GPIB becomes trigger source.
Send
INITiate Initiates the instrument once.
Send
*TRG Triggers the instrument.
Send
FETCh:FREQuency? Fetches the frequency.
Read
<frequency>