COMMAND REFERENCE 4 - 31
ABORt
Syntax:
ABORt
Description:
The ABORt command resets the trigger system and places it in the "IDLE" state.
Pending actions that were already started are finished immediately. The ABORt
command is not finished until the pending actions have been terminated.
Note: The commands
*
RST and ABORt have the same effect on the trigger
functions, except that ABORt does not affect the state of the
INITiate:CONTinuous command. So, when an ABORt command is sent
while the INITiate:CONTinuous is ON, the trigger system will leave the
IDLE state at once.
Example:
Send
ABORt Aborts the current acquisition.
Send
CONFigure:AC Configures for AC-RMS value.
Send
READ:AC? Initiates and reads the AC-RMS value.
Read
<the measured AC-RMS value>