EasyManua.ls Logo

Fluke CombiScope PM3370B - Page 142

Fluke CombiScope PM3370B
297 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4 - 42 COMMAND REFERENCE
Example:
Send
*
RST Resets the instrument.
Send
CALibration Starts auto calibration.
Send
STATus:OPERation:CONDition? Requests for oper. conditions.
Read
<cond_reg> Reads condition register.
WHILE (bit 0 of <cond_reg) = 1) Loops while calibration busy.
Send
STATus:OPERation:CONDition? Requests for oper. conditions.
Read
<cond_reg> Reads condition register.
LOOP_WHILE
Send
STATus:QUEStionable:CONDition? Requests for questionable
conditions.
Read
<cond_reg> Reads condition register.
IF (bit 8 of <cond_reg) = 0)
THEN Calibration_Okay
ELSE (bit 8 of <cond_reg) = 1) Calibration_Not_okay
END_IF
Front panel compliance:
The CALibration command/query is the remote equivalent of the front panel CAL
key.

Table of Contents

Related product manuals