COMMAND REFERENCE 4 - 45
DISPlay:BRIGhtness
Syntax:
DISPlay:BRIGhtness <Numeric_data> | MINimum | MAXimum
<Numeric_data> 0.0 .. 1.0
MINimum Equals 0.0 Trace display is fully blanked.
MAXimum Equals 1.0 Trace display has full intensity.
Query form:
DISPlay:BRIGhtness? [MINimum | MAXimum]
Response:
<NR3>
<NR3> 0.00E00 ... 1.00E00
Description:
The command sets and the query returns the brightness of the trace display. The
number 0.0 (MINimum) gives the lowest brightness. The number 1.0 (MAXimum)
gives the highest brightness.
Notice that the intensity of text display is not controlled with this command.
After a
*
RST command, the brightness is set at 1.80E-01, i.e., 0.18.
Example:
Send
DISPlay:BRIGhtness 0.5 Sets trace brightness at 0.5.
Front panel compliance:
The DISPlay:BRIGhtness command is the remote equivalent of the front panel
TRACE INTENSITY knob.