4 - 54 COMMAND REFERENCE
FETCh?
Syntax:
FETCh[:VOLTage]<measure_function>?
[[ (<voltage_parameters>),] <measure_parameters>]
[,<channel_list> | <trace_list>]
<trace_list> = (@<trace_name>)
<trace_name> = <acquisition_trace> | <memory_trace>
<acquisition_trace> = CH1 | CH2 | CH3 | CH4
These are predefined names for traces
that contain the acquisition result of the
input channels 1 to 4.
Note: CH3 and CH4 not for PM33x0B
<memory_trace> = M<i>_<j>
These are predefined names for traces
that may contain the result of previous
acquisitions or the result of CALCulate
processes.
<i> = Integer value in the range 1 to 50 that
specifies the trace memory register
number.
<i> = 1 to 8: standard memory
<i> = 9 to 50: extended memory
<j> = Integer value in the range 1 to 4 that
specifies the sequence number of the
channel trace in the memory register.
Note: <j>=3 and <j>=4 not for PM33x0B
The other syntax elements are specified with the MEASure? query.
Response:
<NR3)
Example: <1.25E-01> = 0.125