COMMAND REFERENCE 4 - 115
TRIGger[:SEQuence[1]]:FILTer:HPASs:FREQuency
TRIGger[:STARt]:FILTer:HPASs:FREQuency
TRIGger[:SEQuence[1]]:FILTer:HPASs:STATe
TRIGger[:STARt]:FILTer:HPASs:STATe
Syntax:
TRIGger[:SEQuence[1]]:FILTer:HPASs:FREQuency <NRf>
| MINimum | MAXimum
TRIGger[:SEQuence[1]]:FILTer:HPASs:STATe <Boolean>
Alias:
TRIGger[:STARt]:FILTer:HPASs:FREQuency <NRf>
| MINimum | MAXimum
TRIGger[:STARt]:FILTer:HPASs:STATe <Boolean>
<NRf> The cutoff frequency expressed in hertz. The only
possible value is 30000, which defines HF-reject (LF-
pass).
<Boolean> 0 | OFF Sets high-pass filter off.
1 | ON Sets high-pass filter on.
Query form:
TRIGger[:SEQuence[1]]:FILTer:HPASs:FREQuency?
MINimum | MAXimum
Alias:
TRIGger[:STARt]:FILTer:HPASs:FREQuency?
MINimum | MAXimum
Response:
3.00E+04 | 1.00E+08 | 2.00E+08
3.00E+04 Fixed cutoff frequency of 30 KHz (MINimum).
1.00E+08 Bandwidth of 100 MHz
(MAXimum for PM338xB).
2.00E+08 Bandwidth of 200 MHz
(MAXimum for PM339xB).
Query form:
TRIGger[:SEQuence[1]]:FILTer:HPASs:STATe?
Alias:
TRIGger[:SEQuence[1]]:FILTer:HPASs:STATe?
Response:
0 | 1
0 Low-pass filter active.
1 High-pass filter active (HF-reject).