COMMAND REFERENCE 4 - 119
TRIGger[:SEQuence[1]]:HOLDoff
TRIGger[:STARt]:HOLDoff
Syntax:
TRIGger[:SEQuence[1]]:HOLDoff <NRf> | MINimum | MAXimum
Alias:
TRIGger[:STARt]:HOLDoff <NRf> | MINimum | MAXimum
<NRf> The hold-off value expressed in percent.
The range is from 0.00 (MINimum = 0 %) to 1.00
(MAXImum = 100 %).
Query form:
TRIGger[:SEQuence[1]]:HOLDoff? [MINimum | MAXimum]
TRIGger[:STARt]:HOLDoff? [MINimum | MAXimum]
Response:
<NR3>
<NR3> The hold-off value in percent.
Description:
The hold-off value specifies the hold-off time after each Main Time Base sweep,
during which the MTB event detector is inhibited from acting on any new trigger.
For a specification of the minimum and maximum hold-off time, refer to the
Reference Manual supplied. In the digital mode the hold-off time is used to
process previously captured data.
After a
*
RST command, the hold-off value is 0 %.
Example:
Send
TRIGger:HOLDoff 0.5 Hold-off becomes 50 %.
Front panel compliance:
The TRIGger:HOLDoff command is the remote equivalent of the front panel
HOLD OFF knob.