EasyManua.ls Logo

Fluke CombiScope PM3370B - Page 236

Fluke CombiScope PM3370B
297 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
APPLICATION PROGRAM EXAMPLES A - 3
StopEOI% = 256
Termination Receive on EOI
CLS
Clears Output Screen
CALL SendIFC(0)
Clears the GPIB interface
CALL IBTMO(0, 13)
Timeout at 10 seconds
’*** Reset the instrument and clear the status data.
cmd$ = "*RST;*CLS"
CALL Send(0, 8, cmd$, EndEOI%)
CALL errorcheck
’*** Configure for measuring the frequency of the Probe signal.
cmd$ = "CONFigure:VOLTage:FREQuency (0.6),2000,(@1)"
CALL Send(0, 8, cmd$, EndEOI%)
PRINT "Frequency Amplitude Period Pos.width Neg.width"
PRINT " Hertz Volts seconds seconds seconds"
PRINT
’*** Read the signal characteristics 10 times.
FOR i = 1 TO 10
cmd$ = "READ:FREQuency?"
CALL Send(0, 8, cmd$, EndEOI%)
CALL Receive(0, 8, res$, StopEOI%)
Enters frequency
PRINT LEFT$(res$, INSTR(res$, CHR$(10)) - 1),
cmd$ = "FETCh:AMPLitude?"
CALL Send(0, 8, cmd$, EndEOI%)
CALL Receive(0, 8, res$, StopEOI%)
Enters amplitude
PRINT LEFT$(res$, INSTR(res$, CHR$(10)) - 1),
cmd$ = "FETCh:PERiod?"
CALL Send(0, 8, cmd$, EndEOI%)
CALL Receive(0, 8, res$, StopEOI%)
Enters period
PRINT LEFT$(res$, INSTR(res$, CHR$(10)) - 1),
cmd$ = "FETCh:PWIDth?"
CALL Send(0, 8, cmd$, EndEOI%)
CALL Receive(0, 8, res$, StopEOI%)
Enters positive pulse width
PRINT LEFT$(res$, INSTR(res$, CHR$(10)) - 1),
cmd$ = "FETCh:NWIDth?"
CALL Send(0, 8, cmd$, EndEOI%)
CALL Receive(0, 8, res$, StopEOI%)
Enters negative pulse width
PRINT LEFT$(res$, INSTR(res$, CHR$(10)) - 1)
NEXT i
PRINT
CALL errorcheck
END

Table of Contents