Computer Operations
Testing the Instrument/PC RS-232 Interface
4
4-7
Mn.n identifies the main software version.
An.n identifies the analog-to-digital converter software version.
Dn.n identifies the display software version.
Ln.n identifies the programmable gate-array configuration version.
6. Other commands can be entered from the PC to gain familiarity with the instrument
command set. All commands are summarized in Table 4-4 and explained in Table 4-
5.
For example: to reset the instrument, configure channel 0 to measure volts dc using
the 300V DC scale (scale 4), send scan results to the RS-232 port, and scan once,
enter the following:
*RST[Resets the instrument (which does not affect the communication parameters)]
=>
FUNC 0, VDC, 4[Set channel 0 to volts dc and scale 4 (300V DC)]
=>
PRINT_TYPE 0,0[Sets the data destination as the RS-232 port, and all data]
=>
PRINT 1[Enables data logging to the RS-232 port]
=>
*TRG [Triggers a single scan]
=>
15:17:0407/21/94
0:000.00 VDC
ALM:15DIO:255TOTAL:0
To decode the printout, see Figure 5-3.
The commands in the above example can be combined into a single entry by using
the semicolon separator character:
*RST;FUNC 0,VDC,4;PRINT_TYPE 0,0;PRINT 1;*TRG.
7. One of the following three possible prompts are returned when a command is sent to
the instrument:
=> The command was executed[Example, PRINT 1].
!> The command was recognized, but not executed [Example, PRINT 3, where only
PRINT 0 and PRINT 1 are legal entries].
?> The command wasn’t recognized [Example, PRITN 1, spelling error].
8. Exit Windows and return to DOS, as required.
Testing the RS-232 Interface Using Terminal Emulation (Generic)
Complete the procedure below to test the RS-232 link between the PC and instrument
using a generic terminal emulator. Refer to the documentation appropriate to the selected
communications/terminal emulation software, as required. Identify the PC COM port
used for the RS-232 link (COM1 is assumed).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke Hydra Series II and is the answer not in the manual?

Fluke Hydra Series II Specifications

General IconGeneral
BrandFluke
ModelHydra Series II
CategoryTest Equipment
LanguageEnglish

Related product manuals