2635A
Users Manual
4-10
One of the following three possible prompts are returned when a command is sent to
the instrument:
=> The command was executed [Example, PRINT 1].
!> The command was recognized, but not executed [Example, PRINT 3,
where only PRINT 0 and PRINT 1 are legal entries].
?> The command wasn’t recognized [Example, PRITN 1, spelling error].
4. Other commands can be entered from the PC to gain familiarity with the instrument
command set. All commands are summarized in Table 4-4 and explained in Table 4-
5. For example: to reset the instrument, configure channel 0 to measure volts dc
using the 300V DC scale (scale 4), send scan results to the RS-232 port, and scan
once, enter the following [only the output of the last command is shown]:
PRINT #1, "*RST":PRINT INPUT$(10, #1)
PRINT #1, "FUNC 0, VDC,4":PRINT INPUT$(18, #1)
PRINT #1, "PRINT_TYPE 0,0":PRINT INPUT$(20, #1)
PRINT #1, "PRINT 1":PRINT INPUT$(13, #1)
PRINT #1, "*TRG":PRINT INPUT$(83, #1)
15:17:0407/21/94
0:000.00 VDC
ALM:15DIO:255TOTAL:0
To decode the printout, see Figure 5-3.
The commands in the above example can be combined into a single entry by using
the semicolon separator character:
PRINT #1, "*RST;FUNC 0,VDC,4;PRINT_TYPE 0,0;PRINT 1;*TRG":PRINT
INPUT$(124, #1)
If other commands are entered, remember that the input character count xxx for
PRINT INPUT$(xxx, #1) command must be exact. A number too small will not read
all the characters and will leave residual characters in the buffer, while a number too
big will "hang up" the command until more characters are loaded into the buffer or
<CNTL><BREAK> is pressed, which erases the buffer.
5. Enter SYSTEM to exit GWBASIC and return to DOS.
Testing the RS-232 Interface Using Qbasic
Complete the procedure below to test the RS-232 link between the PC and instrument
using QBASIC compiler commands. Identify the PC COM port used for the RS-232 link
(COM1 is assumed).
1. Configure the Data Bucket communication parameters, as described in Figure 4-2,
for bAUd = 9600, PAR = no, CtS = OFF, and Echo = On.
2. Turn on the PC and enter QBASIC to start the BASIC compiler.
3. With the entry screen displayed, enter the following commands (which are not
executed immediately):
OPEN "COM1,9600,N,8,1,CS,CD" FOR RANDOM AS #1
PRINT #1, "*IDN?"

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke Hydra Series II and is the answer not in the manual?

Fluke Hydra Series II Specifications

General IconGeneral
BrandFluke
ModelHydra Series II
CategoryTest Equipment
LanguageEnglish

Related product manuals