Fluke Hydra Series II - Maximum AC Input; Typical Scanning Rate

Fluke Hydra Series II
268 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Appendices
Specifications
A
A-13
Maximum AC Input
300V rms or 424V peak on channels 0, 1, and 11
150V rms or 212V peak on channels 2 to 10 and 12 to 20
Voltage ratings between channels must not be exceeded
2 x 10
6
Volt-Hertz product on any range, normal mode input
1 x 10
6
Volt-Hertz product on any range, common mode input
Cross-Talk Rejection
Refer to Appendix B.
Typical Scanning Rate
See table below. The measurement conditions are: averaged rate over 20 scans;
continuous scanning; alarm limits and Mx+B scanning set on all channels; logging data
to internal memory; and RS-232 communications set at 9600 baud. Measurements were
taken with short-circuit inputs on all channels, except frequency, which was taken with
5V at 15 Hz on all channels.
Table A-17. Typical Scanning Rate
CHANNELS
FUNCTION RANGE SLOW FAST
1 10 20 1 10 20
VDC 300 mV 1.8 3.9 4.1 2.5 13.7 18.4
3V 1.8 3.9 4.1 2.5 13.7 18.4
30V 1.8 3.9 4.1 2.5 13.7 18.3
150/300V 1.8 3.9 4.1 2.5 13.6 18.2
AUTO 1.7 3.6 3.9 2.4 11.3 14.1
TEMPERATURE J (TC) 1.0 3.3 3.8 2.1 12.2 16.6
PT (RTD) 1.7 3.1 3.2 2.1 6.0 6.7
VAC 300 mV 1.1 1.5 1.6 1.3 2.5 2.6
3V 1.1 1.5 1.6 1.3 2.5 2.6
30V 1.1 1.5 1.6 1.3 2.5 2.6
150/300V 1.1 1.5 1.6 1.3 2.5 2.6
AUTO 1.1 1.5 1.5 1.3 2.4 2.5
OHMS 300 1.7 3.1 3.2 2.1 6.0 6.7
3 k 1.7 3.1 3.2 2.1 6.0 6.7
30 k 1.7 3.1 3.2 2.1 6.0 6.7
300 k 1.2 1.8 1.8 1.7 4.0 4.4
3 M 1.2 1.6 1.7 1.7 3.9 4.2
10 M 1.1 1.6 1.6 1.7 3.8 4.0
AUTO 1.7 3.1 3.2 2.1 6.0 6.7
FREQUENCY ANY 0.5 0.6 0.6 0.6 0.7 0.7