EasyManua.ls Logo

Fluke MPS450 - Page 22

Fluke MPS450
36 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MPS450 SERVICE MANUAL CALIBRATION
15
¾ Zero the DMM.
¾ Select NEXT by pressing the [F4] key.
2) ECG High-Range Gain:
¾ Note the words that appear on the LCD screen above the function keys:
DOWN (the [F1] key) decreases the calibration setting.
UP (the [F2] key) increases the calibration setting.
PREV (the [F3] key) returns to the previous calibration step.
NEXT (the [F4] key) advances to the next calibration step.
(The display also indicates the DMM connections, the name of the step, and the required
DMM reading.)
¾ Select UP or DOWN (by pressing the [F2] or [F1] keys) to adjust the MPS450 so that the
DMM reads 5.0 ± 0.005 mV.
¾ Select NEXT by pressing the [F4] key.
¾
3) ECG Low-Range Reference:
¾ Zero the DMM.
¾ Select NEXT by pressing the [F4] key.
4) ECG Low-Range Gain:
¾ Select UP or DOWN (by pressing the [F2] or [F1] keys) to adjust the MPS450 so that the
DMM reads 0.5 ± 0.0005 mV.
¾ Select NEXT by pressing the [F4] key.
5) ECG PACER Gain:
¾ Select UP or DOWN (by pressing the [F2] or [F1] keys) to adjust the MPS450 so that the
DMM reads 10±1 mV.
¾ Select NEXT by pressing the [F4] key.
6) RESPIRATION Reference:
¾ Change the DMM setting to read 30 .
¾ Zero the DMM.
¾ Select NEXT by pressing the [F4] key.
7) RESPIRATION Gain:
¾ Select UP or DOWN (by pressing the [F2] or [F1] keys) to adjust the MPS450 so that the
DMM reads 3 ± 0.03 .
¾ Select NEXT by pressing the [F4] key.

Related product manuals