S IMPLIFIBER ENGLISH - 40
Appendix B - Error Mes-
sages
Err 1 - Invalid Reference value.
Error code 1 indicates that the reference value
is under or over the limits.
- Store a new Reference value.
Err 2 - UNDER or OVER limit.
The measured value is under or over the limit.
The numeric values displayed on the LCD can
range from -55.00 dB/dBm to 3.00 dB/dBm.
When a measurement is less than -55.00 dB/
dBm, the measured value is UNDER the limit.
When a measurement is more than 3.00 dB/
dBm, the measured value is OVER the limit.
- Remeasure
Err 3 - Memory is full, cannot save
result.
Error code 3 indicates that the units memory
is full.